Reference Manual
00809-0100-4340
A
Upgrade Rosemount
to full Oxymitter
Figure A-1: Component Replacement
A. Screws
B. Rosemount OXT adapter board
C. Screws
D. Electronic assembly
Procedure
1. Remove power from the Rosemount Oxymitter DR.
2. Remove the left and right covers from the Rosemount Oxymitter termination
3. Remove and discard screws and Rosemount OXT adapter board (A and B,
4. Place the new Rosemount Oxymitter electronic assembly (D) near the right side of
5. Plug the white connector with the two black wires into the white socket on the
6. Insert the electronics assembly into the termination housing. Ensure the black four-
Rosemount Oxymitter 4000
housing.
A-1) located at the right side of the termination housing.
the termination housing.
bottom power supply card.
wire connector remains outside the housing and in the slot provided in the top card
of the electronics assembly.
The electronics assembly should seat on the bulkhead pins easily. Do not force the
assembly into place.
™
Upgrade Rosemount
Oxymitter DR to full Oxymitter
™
Oxymitter DR
October 2019
Figure
177