Using the Monitor

Setting Patient Adult-Pediatric or Neonatal Modes

60
WARNING: Each time the pulse oximeter is used, check
alarm limits to ensure that they are appropriate for the
patient being monitored.
1. With the monitor in the normal monitoring mode, press the
LIMITS softkey. The monitor displays the ADULT LIMITS or
NEONATE LIMITS screen, depending on the patient setting
used.
NEONATE LIMITS
%SPO2
UPPER
95
LOWER
85
SAT-S
100
SELECT
NEO
2. Press the NEO or ADULT softkey to select ADULT LIMITS or
NEONATE LIMITS depending on the patient being monitored.
%SP02
BPM
190
100
90
BPM
EXIT
ADULT
N-600x Operator's Manual
---
---
NEO