N-600x Operator's Manual
Table 18: Electromagnetic Immunity
-
The N
600x is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
that it is used in an electromagnetic environment as described below.
IEC
Immunity
60601-1-2
Test
Test Level
Power
3 A/m
frequency
(50/60 Hz)
magnetic
field
IEC
-
-
61000
4
8
Note: UT is the AC mains voltage prior to application of
-
600x should assure
Electromagnetic
Compliance
Level
3 A/m
It may be necessary
to position the
-
N
from the sources of
power frequency
magnetic fields or
to install magnetic
shielding. The
power frequency
magnetic field
should be
measured in the
intended
installation location
to assure that it is
sufficiently low.
the test level.