Using OXIMAX Sensor Event Records

Viewing and Printing In-Sensor Tabular Event History Data

92
The ADJUST UP and ADJUST DOWN buttons on the monitor
panel are used to move through the Event Summary table line by
line.
The PRINT softkey enables you to print the displayed event graph.
The BACK softkey returns to the previous TREND/SENSOR
submenu level.
1. With the monitor in the normal monitoring mode, press the
TREND softkey.
2. Press the SENSOR softkey.
3. Press the TABLE softkey to view the data.
SUMMARY
#
DATE
START
100
06JAN
13:55
99
06JAN
11:07
98
06JAN
10:30
97
06JAN
00:02
4. Press the PRINT softkey to print the data.