Covidien Nellcor OxiMax N-600x Operator's Manual page 70

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

Using the Monitor
62
1. Press the LIMITS softkey. The current alarm limits display.
OR
NEONATE LIMITS
%SPO2
UPPER
95
LOWER
85
SAT-S
100
NEO
SELECT
2. Press the ADULT or NEO softkey to select the Adult-Pediatric
or Neonatal alarm limits screen.
3. Press the SELECT softkey to select the parameter to be
adjusted.
4. Use the ADJUST UP or ADJUST DOWN buttons to increase or
decrease the selected limit parameter.
5. Repeat steps 2, 3, and 4 as necessary to complete the alarm
limits setup.
6. Wait for the display to time-out to accept the changes or
press the EXIT softkey to close the display and return to the
normal monitoring mode.
Note: Limit changes are in effect as long as the monitor remains
turned on. When the monitor is turned off, the institutional
default limits are restored. When the monitor is turned on, the
%SP02
BPM
190
100
90
BPM
EXIT
ADULT
N-600x Operator's Manual
---
---
NEO

Hide quick links: