Covidien Nellcor OxiMax N-600x Operator's Manual page 149

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

Time-out
Message
(seconds)
DEFAULTS LOST
None
DELETE
10
TRENDS?
LOW BATTERY
None
1
Pressing the ALARM SILENCE button silences any audible tone and the second press clears the message.
READING
None
TRENDS ...
SENSOR
None
DISCONNECTE
D
N-600x Operator's Manual
Table 13: Prompt/Error Messages
Exit
Exit on
on
Alarm
Alarm
Silence
No
No
Yes
Yes
No
1
Yes
Yes
Yes
No
Yes (1)
Troubleshooting
Displayed
Resolution
-
After the
If the N
600x
monitor is
detects that
-
power
power
on
settings have
been lost.
When
After
attempting to
responding to
delete trend data
the prompt.
from memory by
pressing the
DELETE softkey.
When the
When the
monitor is on
monitor is
battery power
connected to
and the battery
AC power or
charge is low.
when the low
battery is
acknowledged
by pressing the
ALARM SILENCE
button.
When the
When sensor
-
data is
N
600x needs
completely
more than
retrieved or
4 to 6 seconds to
ABORT is
retrieve trend
selected.
data from
memory.
When the sensor
When the
is disconnected
sensor is
from the
reconnected or
monitor.
when the
sensor
disconnection is
acknowledged
by pressing the
ALARM SILENCE
button.
-
cycled.
141