Alarm Limit Changed Indicator

Setting Alarm Limits

N-600x Operator's Manual
Alarm limits that have been changed from the institution or
factory default settings are identified by a decimal point (.) after
the displayed reading (%SpO
is also identified by a decimal point on the alarm limits screen.
The Alarm Limit display enables you to adjust the upper and
lower saturation and pulse rate limits. It also enables you to
adjust the SatSeconds limit.
The Alarm Limit display is accessed by pressing the LIMITS softkey
on the main menu.
The Alarm Limit display includes the alarm limit table and current
measured %SpO
and pulse rate. The title of the alarm limit table
2
indicates whether the instrument is in Adult or Neonate
monitoring mode. If SatSeconds are enabled, the Alarm Limit
display also includes the SatSeconds indicator. Decimal points
after the displayed %SpO
respective limits have been changed from the power-on defaults.
or BPM). The changed parameter
2
or pulse rate indicate that the
2
Using the Monitor
61