Printing Monitor Trend Data

Operating Status

102
Patient information is highlighted in the display above. Parameter
values are displayed directly beneath the heading for each
parameter. In this example, the %SpO
is 190 beats per minute. The "*" next to the 190 indicates that
190 beats per minute is outside of the alarm limits, indicated in
the top row, for pulse rate. If no data for a parameter is available,
three dashes [- - -] display.
PA represents the pulse amplitude value, in which the number
can range from 0 to 254. There are no alarm parameters for this
value. It can be used for trending information as an indication of
a change in pulse volume, relative pulse strength, or circulation.
The Status column indicates alarm conditions and operating
status of the N-600x. In this example, "PH" indicates the pulse
rate upper alarm limit (Pulse High) has been exceeded. A
complete listing of the status codes is listed below. As many as
four codes can be displayed at one time in the Status column.
Code
AO
AS
LB
LM
LP
is 100 and the pulse rate
2
Definition
Alarm Off
Alarm Silence
Low Battery
Loss of Pulse with Interference
Loss of Pulse