Data Port Pinouts - Covidien Nellcor OxiMax N-600x Operator's Manual

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

Using the Data Port

Data Port Pinouts

106
certified to UL-1950 or IEC-60950. The cable used must have a
braided shield that provides 100% coverage, such as a Belden
cable (Belden part number 9609) or equivalent. The shield must
have a 360-degree connection to the metal shell on the N-600x's
DB-15 connector and to the connector on the PC or serial printer.
Caution: Do not create sharp bends in the cable, as this may tear
or break the shielding.
No hardware flow control is used. However, in ASCII mode, XON/
XOFF flow control is supported.
The pinouts for the data port are listed in Table 8.
Table 8: Data Port Pinouts
Pin
1
-
RXD+ (RS
422 [+] input)
2
RXD_232 (RS
3
-
TXD_(RS
232 output)
-
4
TXD+ (RS
422 [+] output)
5
Signal Ground (isolated from Earth Ground)
6
AN_SpO
(analog saturation output)
2
7
NC_NO (relay closure nurse call, normally open)
8
NC_NC (relay closure nurse call, normally closed)
9
-
RxD
(RS_422 [
10
Signal Ground (isolated from Earth Ground)
11
Nurse Call (RS
-
-
12
TxD
(RS
422 [
13
AN_PULSE (analog pulse rate output)
Signal Name
-
232 input)
-
] input)
-
-
-
232
level
output)
-
] output)
N-600x Operator's Manual