Setting SatSeconds Alarm Limit

N-600x Operator's Manual
institutional default limits are effected. Factory or institutional
defaults are selected by qualified service personnel by following
the procedures outlined in the N-600x Service Manual.
Refer to Using SatSeconds on page 153, for more information on
the SatSeconds feature.
Note: The ability to adjust the SatSeconds Alarm limit can be
enabled or disabled by a qualified service personnel as described
in the N-600x Service Manual.
1. With the monitor in the normal monitoring mode, press the
LIMITS softkey. The current alarm limits display.
2. Press the SELECT softkey twice to select %SpO
3. Use the ADJUST UP or ADJUST DOWN buttons to select the
limit. The choices are 10, 25, 50, 100 seconds or OFF.
4. Press the EXIT softkey to save your selection.
Using the Monitor
SAT-S.