Setting Up the Monitor
22
WARNING: Do not lift the pulse oximeter by the pulse
oximetry cable or power cord because the cable or cord
could disconnect from the pulse oximeter, causing the
pulse oximeter to drop on the patient.
WARNING: The N-600x is not defibrillator-proof. However,
it may remain attached to the patient during defibrillation
or while an electrosurgical unit is in use, but the readings
may be inaccurate during the defibrillation and shortly
thereafter.
WARNING: In the USA, do not connect the pulse oximeter
to an electrical outlet controlled by a wall switch, because
the pulse oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable
DOC-10 with the N-600x pulse oximeter. Use of another
pulse oximetry cable will have an adverse effect on
performance. Do not attach any cable that is intended for
computer use to the O
any device other than a Nellcor-approved O
the O
M
sensor connector.
XI
AX
WARNING: The N-600x should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is
necessary, the N-600x should be observed to verify normal
operation in the configuration it is to be used.
Note: The monitor incorporates watchdog timers which reset the
monitor in the event of software errors.