Connecting An Oximax Sensor - Covidien Nellcor OxiMax N-600x Operator's Manual

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

Connecting an O
N-600x Operator's Manual
Note:If the AC power indicator is not lit, check the:
power cord
supply voltage selector switch
user-accessible fuses
AC power outlet
Note: The monitor can be operated with a depleted battery
when connected to an AC power outlet. A warning message
displays and must be cleared by pressing the ALARM SILENCE
button before the monitor can be used for patient monitoring.
M
Sensor
XI
AX
The O
M
sensor type is shown at the bottom of the display
XI
AX
when an O
M
sensor is connected to the N-600x or when the
XI
AX
monitor completes POST with an O
Note: Sensor LED light emissions fall within Class 1 level,
according to IEC 60825-1:2001.
Caution: Use only Nellcor-approved O
pulse oximetry cables.
Note:Physiological conditions, medical procedures, or external
agents that may interfere with the monitor's ability to detect and
display measurements include dysfunctional hemoglobin, arterial
dyes, low perfusion, dark pigment, and externally applied
coloring agents, such as nail polish, dye, or pigmented cream.
Setting Up the Monitor
M
sensor attached.
XI
AX
M
sensors and
XI
AX
25

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents