Using the Monitor
Selecting the Standby Mode
58
WARNING: Do not disable the audible alarm function or
decrease the audible alarm volume if the patient's safety
could be compromised.
1. With the monitor in the normal monitoring mode, press the
ALARM SILENCE button until the alarm silence duration
setting displays.
2. While pressing the ALARM SILENCE button, press and hold
the ADJUST UP button until OFF displays. Release the buttons.
The standby mode enables the monitor to retain the alarm limit
settings that are in effect while monitoring a patient. The monitor
must be connected to an AC power source to enter the standby
mode.
Normally, the standby mode setting is used when a patient has to
temporarily leave the monitor.
N-600x Operator's Manual