154
The alarm response time, assuming a SatSeconds limit set at 50
and a lower alarm limit set at 90, is described and illustrated
below.
In this example, the %SpO
remains there for a period of 2 seconds (2 points x 2 seconds = 4
SatSeconds). The %SpO
then to 84 for 6 seconds. The resulting SatSeconds are:
%SpO
2
2 x
4 x
6 x
Total SatSeconds =
After approximately 10.9 seconds the SatSeconds alarm would
sound, because 50 SatSeconds had been exceeded. See the
arrow () in Figure 11.
Figure 11: Alarm Response with SatSeconds
level drops to 88 (2 points) and
2
then drops to 86 for 3 seconds and
2
Seconds
SatSeconds
2 =
4
3 =
12
6 =
36
52
N-600x Operator's Manual