Emi (Electromagnetic Interference) - Covidien Nellcor OxiMax N-600x Operator's Manual

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

Table 14: Common Problems and Resolutions
Problem
Error Code: "EEE XXX"
followed by a number
appears.

EMI (Electromagnetic Interference)

N-600x Operator's Manual
Caution: This device has been tested and found to comply
with the limits for medical devices to the IEC 60601-1-1-2
(second edition), EN60601-1-2, and the Medical Device
Resolution
Press the ON/STANDBY button and allow
the monitor to shut off completely. Then
press the button again to turn the
monitor back on.
If the error code persists, record the
number and provide this information to
a qualified service personnel, or your
local Nellcor representative.
Error Code "EEE 529 or 729" displays
when the battery discharges to a
critically low level.
Ensure the SUPPLY VOLTAGE SELECTOR
switch on the rear panel is set to the
proper voltage based on your location.
Press the ON/STANDBY button and allow
the monitor to shut off completely.
Allow the battery to charge for about 10
minutes and then turn the unit back on.
If the error code is still present, turn the
unit off and allow it to continue to
charge. If the monitor has been charged
for 30 minutes and the error code
persists, notify a qualified service
personnel, or your local Nellcor
representative.