N-600x Operator's Manual
the screen and display the message "EEE XXX," and sound a low
priority alarm. Cycling the power clears these errors.
Table 12: Error Codes
Error
Error
Code
Message
80
DEFAULTS
LOST
81
SETTINGS LOST
82
CLOCK
SETTING LOST
515, 518,
N-600x Boot
534, 535,
Version x.x.x.x
569
529, 729
LOW BATTERY
Troubleshooting
Action
-
The current power
on default
settings have been lost and returned
to factory defaults. Qualified service
N-600x
personnel can use the
Service Manual
to restore the
-
desired power
on default settings.
The current settings (for example,
alarm limits, alarm and pulse beep
volumes, alarm silence duration) have
been lost and returned to power
defaults. Turn the monitor off and
back on again. If it is necessary to
have settings different from the
-
power
on default settings, turn the
monitor off and back on again, and
reenter the desired settings.
The date and time settings have been
lost. Reenter the date and time.
Recharge or replace the battery.
The application software is missing
or corrupt. Notify a qualified service
personnel.
The battery is discharged to a critically
low level. The monitor will shutdown
after 10 seconds.
Verify the SUPPLY VOLTAGE
SELECTOR switch on the rear panel is
set to the proper voltage.
Connect the monitor to AC power
and turn back on. A warning
message is displayed and a low
priority audible alarm sounded. These
must be acknowledged (by pressing
the alarm silence key twice) before
the monitor can be used for patient
monitoring.
-
on
137