Operating the Battery
Critical
Low
AC
State
Operation
Battery
Battery
Power
7
Yes
Yes
Yes
SpO2 - displayed.
AC/Battery Charge LED -
on
LOW BATTERY LED-on
(flashing)
LOW BATTERY message -
on
The Battery Fuel Gauge
Indicator shows a fully
depleted battery (no bars
lit).
Warning message in the
pleth window: UNIT WILL
SHUT DOWN IF AC
POWER LOST
Audio alarm - low priority
Error code - logged
Affect of Silence key - One
press silences the audible
alarm. Pressing the Affect
Silence key twice cancels
the LOW BATTERY
message, removes the
warning message and
restores default Pleth (or
Blip) display (LED
continues to FLASH until
Low Battery condition is
not true, Battery Fuel
Gauge Indicator shows
charging progress)
Shutdown - N/A
N-600x Operator's Manual
33