Covidien Nellcor OxiMax N-600x Operator's Manual page 183

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

N-600x Operator's Manual
Table 18: Electromagnetic Immunity
-
The N
600x is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
that it is used in an electromagnetic environment as described below.
IEC
Immunity
60601-1-2
Test
Test Level
Electrostatic
±6 kV contact
discharge
(ESD)
±8 kV air
IEC
-
-
61000
4
2
Electric fast
±2 kV for
transient/
power supply
burst
lines
IEC
±1 kV for
-
-
input/output
61000
4
4
lines
Surge
±1 kV
differential
mode
IEC
-
-
61000
4
5
±2 kV
common
mode
Note: UT is the AC mains voltage prior to application of
-
600x should assure
Electromagnetic
Compliance
Level
±6 kV contact
Floor should be
wood, concrete, or
ceramic tile. If
floors are covered
±8 kV air
with synthetic
material, the
relative humidity
should be at least
30%.
±2 kV for
Mains power
power supply
quality should be
lines
that of a typical
commercial and/or
hospital
environment.
±1 kV for
input/output
lines
±1 kV
Mains power
differential
quality should be
mode
that of a typical
commercial and/or
hospital
environment.
±2 kV
common
mode
the test level.
Specifications
Environment
Guidance
175