About The Audible Indicators - Covidien Nellcor OxiMax N-600x Operator's Manual

Hide thumbs Also See for Nellcor OxiMax N-600x:
Table of Contents

Advertisement

Indicator
Data In-Sensor
Indicator
SatSeconds Indicator
Fast Response Mode
Indicator
Neonate Alarm Limits
Indicator

About the Audible Indicators

N-600x Operator's Manual
Table 2: Visual Indicators
Lights to indicate that the attached O
patient sensor event record. The sensor event record
information may be viewed or printed.
Fills in clockwise as the SatSeconds alarm management system
detects a %SpO
reading outside of the limit setting. Empties in
2
counterclockwise direction when %SpO
limits. When the indicator is full, a medium priority alarm
sounds.
Determines the response time (2 to 4 seconds in fast mode and
5 to 7 seconds in normal mode) applied by the O
algorithm in its calculation of SpO
calculation of pulse rate is unaffected by the response mode
setting. The trending interval (2 seconds or 4 seconds) updated
automatically by the monitor to roughly correspond with the
SpO
calculation response time.
2
Displays when the alarm limits are set to neonate. No symbol
displays when the monitor is set to adult limits.
Table 3: Audible Indicators
Function
Alarm Silence
Reminder
Confirmation Tone
Symbols, Controls, Displays and Indicators
Description
M
sensor contains a
XI
AX
reading is within
2
. The O
M
2
XI
AX
Description
Three beeps sound approximately every
three minutes when alarms are silenced with
the alarm silence duration set to OFF and
the alarm silence reminder function is
enabled.
Three beeps sound to indicate default
settings have been saved or reset to factory
defaults or trend data has been deleted.
M
XI
AX
algorithm's
17