138
Table 12: Error Codes
Error
Error
Code
Message
575
TRENDS LOST
701-716,
POWER SUPPLY
720-724,
FAILURE
732-740,
576-582
717, 718
BATTERY
FAILURE
725-728,
REPLACE
730
BATTERY
Action
Monitor trends are corrupted and will
be cleared.
Turn off the monitor, and then back
on again.
The monitor power supply has
detected an error. The monitor will
shutdown after 10 seconds.
Verify the monitor is being operated
within specified environmental
conditions. Notify a qualified service
technician.
The monitor has detected a battery
open or short condition. The monitor
will shutdown after 10 seconds.
Battery should be replaced. Notify a
qualified service technician.
The battery is not charging properly.
The monitor will shutdown after
10 seconds.
Battery should be replaced. Notify a
qualified service technician.