About the Visual Indicators

Indicator
%SpO
Display
2
Pulse Amplitude
Indicator (blip bar)
Plethysmographic
Waveform Display
Pulse Rate Display
AC Power Indicator
Low Battery Indicator
N-600x Operator's Manual
Table 2: Visual Indicators
Shows the hemoglobin oxygen saturation level. The display
value flashes zeros during loss-of-pulse alarms and flashes the
SpO
value when the SpO
2
Pulse Search, the monitor continues to update the display. If
alarm limits have been changed from their power-on defaults, a
decimal point (.) is displayed after the SpO
Indicates pulse beat and shows the relative (non-normalized)
pulse amplitude. As the detected pulse becomes stronger, more
bars light with each pulse. This indicator is available only in the
blip view.
Displays a non-normalized waveform in real-time sensor signals.
The relative pulsatile strength and quality of the incoming
signals can be observed.
Displays the pulse rate in beats per minute. It flashes during
loss-of-pulse alarms and when the pulse rate is outside of the
alarm limits. During Pulse Search, the monitor continues to
update the display. Pulse rates outside of the pulse rate range of
20 to 250 bpm are displayed as 0 and 250, respectively. If alarm
limits have been changed from their power-on defaults, a
decimal point (.) is displayed after the BPM value (112.).
Lights continuously when the N-600x is connected to an AC
power source. The indicator shows that the battery is charging.
It is off when the monitor is being powered by internal battery.
Lights continuously when 15 or fewer minutes of battery
capacity remain. Flashes when the battery capacity reaches a
critically low condition.
Symbols, Controls, Displays and Indicators
Description
is outside the alarm limits. During
2
value (81.).
2