Table 2-11.  Patient Applied And Patient Isolation Risk Current - Covidien Nellcor GR101704 Service Manual

Bedside respiratory patient monitoring system
Table of Contents

Advertisement

Table 2-10.  Earth and Enclosure Leakage Current Specifications (Continued)
Condition
Normal
Single Fault
Normal
Single Fault
Condition
Normal
Single Fault
Normal
Single Fault
Condition
Single Fault
Service Manual
Enclosure Leakage Current
Neutral
AC Line Polarity
Line Cord
Normal
Closed
Open
Closed
Reversed
Closed
Open
Closed
Table 2-11.  Patient Applied and Patient Isolation Risk Current
Patient Applied Risk Current
AC Line Polarity
Neutral Line
Normal
Closed
Open
Closed
Reversed
Closed
Open
Closed
Patient Isolation Risk Current
AC Line Polarity
Neutral Line
Normal
Closed
Reversed
Closed
Manufacturer's Declaration and Guidance
Power Line
Ground
ANSI/AAMI 60601-1
Closed
Closed
Open
Closed
Closed
Open
Power Line
Ground Cable
ANSI/AAMI 60601-1
Closed
Closed
Open
Closed
Closed
Open
Power Line
Ground Cable
Closed
Closed
IEC 60601-1
100 μA
500 μA
100 μA
500 μA
IEC 60601-1
100 μA
500 μA
100 μA
500 μA
IEC 60601-1
UL 60601-1
5000 μA
2-13