Using the Data Port

Overview

Connecting to the Data Port

N-600x Operator's Manual
Patient data can be output through the data port on the back of
the N-600x by connecting it to a PC or serial printer.
When connecting the N-600x to a printer or PC, verify proper
operation prior to clinical use. Both the N-600x and the printer or
PC must be connected to a grounded AC power outlet. The
N-600x protocol setting must be ASCII.
Any printer or PC connected to the monitor's data port must be
certified according to IEC Standard 950. All combinations of
equipment must be in compliance with IEC Standard 60601-1-1
systems requirements. Anyone who connects a printer or PC to
the data output port configures a medical system and is therefore
responsible for ensuring the system complies with the
requirements of system standard IEC Standard 60601-1-1 and the
electromagnetic compatibility system standard IEC Standard
60601-1-2.
The N-600x data port may be connected to a serial printer or PC
by using a cable terminated with:
an AMP connector (AMP part number 747538-1),
ferrule (AMP part number 1-747579-2), and
compatible pins (AMP part number 66570-2).
The cable should not exceed 25 feet (7.6 meters) in length. The
external ITE (Information Technology Equipment) device must be
Using the Data Port
105