Performance Considerations
128
failure to cover the O
material in high ambient light conditions
Loss-of-pulse signal can occur for the following reasons:
the O
M
XI
AX
a blood pressure cuff is inflated on the same extremity
as the one with the O
there is arterial occlusion proximal to the O
sensor
poor peripheral profusion
Select an appropriate O
observe all warnings and cautions presented in the directions for
use accompanying the O
substances such as nail polish from the application site.
Periodically check to ensure that the O
properly positioned on the patient.
High ambient light sources such as surgical lights (especially those
with a xenon light source), bilirubin lamps, fluorescent lights,
infrared heating lamps, and direct sunlight can interfere with the
performance of an SpO
from ambient light, ensure that the sensor is properly applied,
and cover the O
M
sensor site with opaque material.
XI
AX
WARNING: Failure to cover the O
opaque material in high ambient light conditions may
result in inaccurate measurements.
If patient movement presents a problem, try one or more of the
following remedies to correct the problem.
verify the O
applied
move the O
M
sensor site with opaque
XI
AX
sensor is applied too tightly
M
sensor attached
XI
AX
M
sensor, apply it as directed, and
XI
AX
M
sensor. Clean and remove any
XI
AX
M
sensor remains
XI
AX
O
M
sensor. To prevent interference
2
XI
AX
M
sensor site with
XI
AX
M
sensor is properly and securely
XI
AX
M
sensor to a less active site
XI
AX
N-600x Operator's Manual
M
XI
AX