O
M
Sensor Event Record Graphical Data
XI
AX
N-600x Operator's Manual
sensor memory chip in the prior monitoring situation, a TREND/
SENSOR option is unavailable.
A sample event display without data is shown below. The
message clears when the graph or summary closes.
Graphical representations of patient event history is only available
on single-patient-use O
the minimum or maximum %SpO
interval throughout the duration of an event (%SpO
continuously below alarm threshold for at least 15 seconds) and
continuing every 30 seconds until the actual %SpO
or exceeds the alarm threshold.
The duration of an event is determined by the number of data
points in the event. Each data point is stored at 30-second
intervals.
Events end due to one of the following reasons:
%SpO
returns to or above the alarm limit
2
Loss of pulse
Using OXIMAX Sensor Event Records
M
sensors. Graphed data points are
XI
AX
value for each 30-second
2
2
value equals
2
87