Using the Monitor

Setting Monitor Response Mode

64
The purpose of the response mode is to set the response time of
the N-600x algorithm calculation of the SpO
does not affect the N-600x algorithm's calculation of pulse rate).
The trending interval (2- or 4-seconds) is updated automatically
by the monitor to roughly correspond with the SpO
response time.
The response mode automatically programs the N-600x algorithm
to record and display monitor trend information at 2-second
intervals (Fast Mode) up to 24 hours or 4-second intervals
(Normal Mode) up to 48 hours.
The response mode display screen includes the current SpO
response mode setting and the current measured %SpO
pulse rate. When in the fast mode, the screen displays the fast
mode symbol.
1. With the monitor in the normal monitoring mode, press the
SETUP softkey.
2. Press the NEXT softkey three times.
3. Press the MODE softkey.
Note: When the monitor is in the fast response mode, the
monitor may produce more SpO
are accustomed to experiencing.
(the response mode
2
calculation
2
and pulse rate alarms than you
2