Adult-Pediatric or Neonatal Settings

N-600x Operator's Manual
1. Verify the N-600x is monitoring a patient and the alarm limits
are configured to the patient being monitored.
2. Disconnect the sensor from the monitor.
3. Press the ALARM SILENCE button to silence the audible
alarms.
4. Press the ALARM SILENCE button to disable the alarm
messages.
The monitor is now in the standby mode. Reconnect the sensor
to the monitor and the patient to return to normal monitoring.
The clinician can set the monitor's operating mode to
adult-pediatric or neonatal by using the LIMITS softkey. This
setting remains active until the monitor is turned OFF. The factory
default power-on setting is for adult-pediatric patients. This
default setting can be changed to neonatal by qualified service
personnel using the procedures indicated in the N-600x Service
Manual.
Refer to Table 15, for neonate factory default limit settings. Refer
to Table 16, for adult factory default limit settings.
Using the Monitor
59