An instance of mixed dosing (manual and automatic) has been extended by
signalling its corresponding phases to present the possibilities of the working
mode <Dosing>.

28.7. Reporting of completed dosing processes

On completing each dosing process, the terminal automatically generates
a report on that process.
Caution:
Submenu: "
report printout template" enables optional modifying of the report
template (see ch. 16.2.3 of this user manual).
Value
Signal flag 1 – „1"
Tare
[5s] Open valve
Water
2 kg [Water]
[5s] Close valve
Water
Output 12 – „1"
Signal flag 1 – „1"
Signal flag 2 – „1"
[10s] Dosing
completed
Devices /
Printer /
Description
Checking condition, whether signal flag 1 is set to
value "1" – i.e. checking whether desired part of the
process has been already realized on the weighing
platform no. 1. If Yes, then dosing process on the
weighing platform no. 2 shall start.
Tarring the weighing platform no. 2
Pending for opening main valve of the container with
"Water"
Automatic dosing of ingredient "Water" to obtain 2 kg
mass in single-step dosing mode – Output 6 controls
the dosing valve (set for Water in Products database)
Pending for closing the main valve of the container
"Water"
Output 12 takes high status ("1") – signal confirming
carrying out the dosing process on the weighing
platform no. 2 is lit up
Setting the characteristic point of the process, which is
a condition for carrying out part of the process on the
weighing platform no. 1
Displaying a text message on the terminal confirming
completion of dosing process
Printouts /
140
Dosing