14.2. Filter
This filter is intended to suppress continuous mechanical vibrations at the
cost of stabilization time.
Procedure:
• Enter <
<
Filter> and then set an appropriate value.
Accessible settings:
None, V.Fast, Fast, Average, Slow.
Caution:
The higher filtering level the longer stabilization time.
14.3. Autozero
The autozero function has been implemented in order to assure precise
indications. This function controls and corrects „0" indication.
While the function is active it compares the results continuously with
constant frequency. If two sequential results differ less than the declared
value of autozero range, so the scale will be automatically zeroed and the
pictograms
If AUTOZERO is disabled zero is not corrected automatically. However,
in particular cases, this function can disrupt the measurement process e.g.
slow pouring of liquid or powder on the weighing pan. In this case, it is
advisable to disable the autozero function.
Procedure:
• Enter <
<
Autozero> and then set an appropriate value.
Accessible settings:
Weighing> according to ch. 14 of this manual, select
and
will be displayed.
Weighing> according to ch. 14 of this manual, select
-
Autozero off
-
Autozero on
48