Ive Balance Serial Number; Et Mass Value Of A Single Item; Et Reference Mass Value - RADWAG WLY User Manual

Y series
Hide thumbs Also See for WLY:
Table of Contents

Advertisement

37.4.20. Give balance serial number
Syntax:
NB CR LF
Possible answers:
NB_A_"x" CR LF
NB_I CR LF
x – serial number of the device (inserted in between inverted commas)
Example:
NB CR LF – give serial number
NB_A_"123456" CR LF – serial number of the device – 123456
37.4.21. Set mass value of a single item
Syntax: SM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Possible answers:
SM_OK CR LF
SM_I CR LF
ES CR LF
Notice:
Command supported for „Parts counting" mode exclusively.
37.4.22. Set reference mass value
Syntax: RM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Possible answers:
RM_OK CR LF
RM_I CR LF
ES CR LF
Notice:
Command supported for „Deviations" mode exclusively.
- command understood, returns serial number
- command understood but not accessible at this moment
- command understood
- command understood but not accessible at this moment
(e.g. mode other than „Parts counting")
- command not recognised (mass format incorrect)
- command understood
- command understood but not accessible at this moment
(e.g. mode other than „Deviations")
- command not recognised (mass format incorrect)