Send The Stable Result In Basic Unit; End The Result Immediately In Basic Unit - RADWAG WLY User Manual

Y series
Hide thumbs Also See for WLY:
Table of Contents

Advertisement

Caution:
Use dots as decimal points in tare values.

37.4.5. Send the stable result in basic unit

Syntax: S CR LF
Possible answers:
S_A CR LF
S_E CR LF
S_I CR LF
S_A CR LF
MASS FRAME
Frame format:
1
2-3
S
space
Example:
S CR LF – computer command
S _ A CR LF - command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.
37.4.6. Send the result immediately in basic unit
Syntax: SI CR LF
Possible answers:
SI_I CR LF
MASS FRAME
Frame format:
1
2
3
S
I
space
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in basic unit is returned
4
5
stability
space
- command comprehended but cannot be executed at the moment
- mass value in basic unit is returned
4
5
stability
space
6
7-15
16
sign
mass
space
6
7-15
sign
mass
space
246
17
18
19
unit
16
17
18
unit
20
21
CR
LF
19
20
21
CR
LF

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wpy

Table of Contents