37.4.17. Read upper threshold
Syntax: OUH CR LF
Possible answers: UH_MASA CR LF - command executed
Frame format:
1
2
U
H
Mass
Unit
37.4.18. Simulation of pressing ENTER/PRINT key
Syntax: SS CR LF
Sending a command S S CR LF to a scale causes automatic saving of a
weighing record in the database and simultaneous activating a declated
printout template.
Caution:
While sending the command to a scale, all criteria for successful
measurement execution have to be met, i.e. result control, stable mass
indication, etc.
37.4.19. Change platform
Syntax: PN CR LF, where N – number of platforms (1 to 4)
Possible answers:
PN_OK CR LF
PN_I CR LF
ES CR LF
3
4-12
space
mass
- 9 characters justified to the right
- 3 characters justified to the left
- command completed
- command correct, but not accessible at the moment
- command incorrect (e.g. incorrect number of platforms)
13
14
15
space
unit
251
16
17
18
space
CR
19
LF