Omron CJ1W-SCU32 Operation Manual page 450

Machine automation controller, cj / nj-series, serial communications units, cpu unit
Table of Contents

Advertisement

A Standard System Protocol
High Calibration Set (Sequence No. 446 (Hex 01BE))
This sequence sets the Laser Micrometer's high calibration.
Send Data Word Allocation (3rd Operand of PMCR(260))
The allocation is similar to that of sequence No. 404 (Calibration Set) except that the LC gauge
dimension in words +4 to +6 are not used.
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
Additional Information
• Use sequences No. 446 (High calibration set) and No. 447 (Low calibration set) to calibrate the
Laser Micrometer. If sequence No. 404 (Calibration Set) is used, both the high and low calibra-
tion gauges must be set, and the high calibration and low calibration gauges cannot be
exchanged.
• The HC gauge dimension must be set with 3 digits for the integer portion and 4 digits for the
decimal portion.
Low Calibration Set (Sequence No. 447 (Hex 01BF))
This sequence sets the Laser Micrometer's low calibration.
Send Data Word Allocation (3rd Operand of PMCR(260))
The allocation is similar to that of sequence No. 404 (Calibration Set) except that the HC gauge
dimension in words +1 to +3 are not used.
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
Additional Information
• Use sequences No. 446 (High calibration set) and No. 447 (Low calibration set) to calibrate the
Laser Micrometer. If sequence No. 404 (Calibration Set) is used, both the high and low calibra-
tion gauges must be set, and the high calibration and low calibration gauges cannot be
exchanged.
• The LC gauge dimension must be set with 3 digits for the integer portion and 4 digits for the
decimal portion.
R-116
CJ-series Serial Communications Units Operation Manual for NJ-series CPU Unit(W494)