PRINTER SYMPTOM
Printer is feeding paper, but nothing is printed.
Printer is not printing and POWER indicator is
red.
Printer POWER indicator does not illuminate
when printer is turned on.
Printer is not printing and STATUS indicator is
orange.
Printer is not printing and STATUS indicator is
red.
CLEANING THE i-STAT PRINTER
Clean the external casing of the i-STAT Printer with a gauze pad moistened with any of the following ap-
proved cleaning agents:
10% bleach solution
Isopropyl alcohol (IPA)
PDI
Super Sani-Cloth
®
chloride)
Rinse the printer casing using another gauze pad moistened with water and dry.
DO NOT IMMERSE THE PRINTER IN ANY FLUID, AT ANY TIME.
7 - 16
Check that the paper is feeding from under the
roll.
Battery needs to be recharged.
Battery needs to be recharged.
Printer is out of paper.
Print head temperature is out of range. Print-
ing will be inhibited until print head temperature
returns to normal level.
(solution of IPA, n-Alkyl dimethyl ethylbenzyl- and benzyl- ammonium
®
Art: 714369-01K
RECOMMMENDED ACTION(S)