Method comparison data were collected using a modification of the CLSI guideline EP9-A.
arterial blood samples were collected in plastic syringes and analyzed in duplicate on the i-STAT System
and in duplicate using the comparative methods. All samples were analyzed immediately upon collection.
The patient populations in the studies were those in which ACT is routinely used. This includes baseline,
heparin-treated, and heparin-reversed samples from patients undergoing cardiac catheterization and
cardiac bypass.
Deming regression analysis
table, n is the number of specimens in the data set, Sxx and Syy refer to estimates of the imprecision based
on the duplicates of the comparative and i-STAT methods respectively, Sy.x is the standard error of the
estimate, and r is the correlation coefficient.
Method comparisons will vary from site to site due to differences in the sample handling, reagent and
instrument systems in use, and other site-specific variables.
Cath Lab
Medtronic
HR-ACT
n
270
Sxx
10.1
Syy
10.7
Slope
1.15
Int't
-30
Sy.x
32.5
Xmin
73
Xmax
523
r
0.848
Rev. Date: 01-Jul-13
was performed on the first replicate of each sample. In the method comparison
3
Hemochron
CA510/FT CA510
418
19.7
13.5
0.86
-3
22.5
63
763
0.903
Art: 714185-01P
Hemchron CA510/FT CA510
CVOR
Site 1
Site 2
n
35
Sxx
15.8
34.2
Syy
13.0
11.5
Slope
0.85
1.10
Int't
4
Sy.x
43.8
17.4
Xmin
118
Xmax
671
735
r
0.912
0.952
Venous or
2
Site 3
30
24
24.4
20.8
1.19
-52
-73
62.1
94
125
767
0.891
Celite ACT - 3