Abbott i-STAT 1 System Manual page 314

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

Configure for
Channel 2 provides network access for the i-STAT 1 Analyzer data transmissions
i-STAT 1 Data
to a data manager. This section describes how to set up parameters for Channel
Transfer via IR Port
2.
1. Determine the following information:
2. At the Your choice ? prompt, Select 2 (Channel 2 Configuration).
3. At each of the prompts enter the following bold-faced value:
Verify and Save
1. When the Summary screen appears, verify that the information you
Settings
21-4
The IP Address of the data manager. (Example: 10.10.12.184)
The service port number set to receive transmissions from the
i-STAT 1 Analyzer (default 6004).
Note: If the information to be entered is the same as the default
value, press the Enter key.
Baudrate (38400) ? 38400 (must be set to 38400)
I/F Mode (4C) ? (press <Enter> key)
Flow (00) ? (press <Enter> key)
Port No (10002) ? (press <Enter> key)
ConnectMode (C1) ? C1 (must be set to C1)
Auto increment source port (N) ? N (press <Enter> key)
Remote IP Address : (000)10.(000)10.(000)12.(000) 184
Note: Set Remote IP Address to the IP Address of the
computer where the data manager resides.
Remote Port (06004) ? 6004
Note: Remote Port refers to i-STAT 1 service port defined in
the data manager.
DisConnMode (00) ? (press <Enter> key)
FlushMode (44) ? 44 (must be set to "44")
DisConnTime (00:30) ? 00:30 (Disconnect time must be 30
seconds)
SendChar 1 (00) ? (press <Enter> key)
SendChar 2 (00) ? (press <Enter> key)
entered is correct. If it is not, fix the appropriate settings and continue.
Art: 714383-01F
Rev. Date: 02-Sep-08

Advertisement

Table of Contents
loading

Table of Contents