ANALYZER CODED MESSAGES
From the time it powers up until the time it powers down, the i-STAT
checks. The failure of any quality check causes the analyzer to halt the test cycle and display a "cause", an
"action" message, and a code.
The Cause Message:
This message describes the likely cause of the failed quality check. For example, when an overfilled
cartridge is detected, the analyzer will display "Sample Positioned Beyond Fill Mark".
The Action Message:
This message indicates the appropriate action. For example, if it is likely the quality check will fail again the
next time the analyzer is used, the instruction "Use Electronic Simulator" will be displayed. If the problem is
related to an operator or cartridge, the instruction "Use Another Cartridge" will be displayed.
The Cause Code:
This is a numeric code associated with the failed quality check. Since multiple codes can be associated
with a single cause message, this is essential information when contacting your local support organization
for further assistance. The codes are stored in the analyzer's memory along with other test records and are
transmitted to the Central Data Station. The code list can be viewed and printed.
Codes 1-15 and 95 usually indicate a condition related to the environment or the state of the analyzer.
These conditions are usually benign and go away after the next cartridge or Electronic Simulator is
inserted, or after the offending condition is corrected.
Code
Cause/Action
Number
Message on Display
1
Dead Batteries /
Replace Batteries
Abbott Point of Care Inc. • 100 & 200 Abbott Park Road • Abbott Park, IL 60064 • USA
i-STAT
TECHNICAL BULLETIN
There is insufficient battery power to complete the testing
cycle. Replace the disposable lithium batteries in the analyzer
or recharge the rechargeable batteries.
If you are experiencing this code frequently and use disposable
batteries with the i-STAT 1 Analyzer, you may want to consider
the rechargeable battery system available with the i-STAT 1
Analyzer.
Art: 714260-01R
®
Analyzer performs numerous quality
®
Explanation
Rev. Date: 30-Mar-17