TRANSMITTING DATA FROM THE DRC TO THE DATA MANAGER
1. Place handheld in the DRC's cradle. If properly aligned, the blue proximity light will turn on and
a "Waiting to Send" message will be displayed on the handheld until communication is estab-
lished with the Data Manager.
2. Once the handheld establishes communication with the Data Manager, a "Communication in
Progress" message will then appear on the handheld display and the arrows will circle until the
transmission is complete.
Note: Do not move handheld until the "Communication in Progress" message disappears.
TRANSMITTED INFORMATION
The following information is transmitted from the i-STAT 1 handheld with each test record:
•
Date and time the test was performed
•
Operator and Patient ID or Quality Test fluid lot number
•
All information entered by the operator, e.g. lot numbers, sample types, and comment codes
•
Result(s)
•
Serial number of the handheld
•
Uses count on the handheld
•
Application software version in the handheld
•
CLEW standardization software in the handheld
CHARGING BATTERIES BEFORE USE
Place a new rechargeable battery in the recharging compartment on the DRC for 40 hours. The battery
will then be 100% charged and ready for use. A handheld with disposable batteries may be placed on
the DRC to download data until the rechargeable battery is ready.
RECHARGEABLE BATTERY LIFE
A fully charged battery, if not periodically recharged, will self-discharge in approximately 3 months.
Prevent self-discharge of the battery by either:
•
Keeping the rechargeable battery in a handheld that is periodically placed on the DRC, or
•
Storing the rechargeable battery separately in the external charging bay on the DRC.
Rev. Date: 14-Mar-14
Art: 728690-01F
5