Sample Handling
System
Contact
Pads
Sensors
Sample Well
Sensors
Contact Pads
Heating Elements
3-2
Part
Function
Sensor Channel
The sensor channel directs the sample from the sample
chamber to the sensors. An extension of this channel
becomes a waste chamber to receive the calibrant
solution, if applicable, as it is displaced by the sample.
Air Chamber
An air chamber is located in blood gas/electrolyte/
chemistry/hematocrit cartridges between the sample
chamber and sensor channel. This creates an air
segment between the calibrant solution, if applicable,
and the sample to prevent the two from mixing. The
size of the air segment is monitored by the analyzer.
Sample Chamber
The sample chamber includes the sample well and the
channel leading from the well up to the fill mark. When
filled, the sample chamber contains sufficient sample for
testing. Sample volume and placement are monitored
by the analyzer.
Bladder
The bladder (concealed by the label) is connected to
the sample well. The analyzer presses on the bladder
to displace calibrant solution, if applicable, from the
sensors to move the sample from the sample chamber
to the sensors or to mix sample and reagents.
Snap Closure
The snap closure creates an airtight seal necessary for
proper fluid movement within the cartridge. The closure
(if applicable)
also ensures that calibrant, if applicable, and sample
remain contained within the cartridge during the testing
cycle and subsequent disposal. Immunoassay cartridges,
such as cTnI, use a plastic slide enclosure clip.
Air Vent
An air vent on the underside of the cartridge, beyond
the fluid front, allows the calibrant and the sample to
flow forward, but not out of the cartridge.
Waste Chamber
A waste chamber (beneath the cartridge label) holds
calibrant fluid after it has been used.
The sensors are electrodes microfabricated on silicon chips. Electrodes have
chemically sensitive coatings such as ion-selective membranes and enzyme
layers. In cartridges that perform coagulation tests, reagents, such as clot
activators, are coated on the plastic above the sensors. Each sensor is connected
to a contact pad by a signal line. The sensors respond to the calibrant solution,
if applicable, and the sample by producing measurable signals related to
analyte concentration. The performance characteristics for each sensor are
described in the Cartridge and Test Information section. The section on theory
describes the measurement principles.
The contact pads conduct the signals generated by the sensors to the analyzer.
In order to function properly, care must be exercised not to contaminate the
contact pads during cartridge handling.
All i-STAT cartridges require thermal control at 37 °C, and include heating
elements on the underside of the sensor chips which are contacted and heated
by the handheld's thermal probes.
Art: 714365-01G
Rev. Date: 11-Mar-14