Target values are specific to the i-STAT System. Results may differ if used with other methods.
If a result for a level is outside the range published in the Value Assignment Sheet, two additional cartridge
runs should be performed on this level and the three results averaged and then compared to the Value
Assignment Sheet range. If this average value is still outside the acceptable range, see the Troubleshooting
section below for additional procedures to follow.

PROCEDURE FOR CARTRIDGES

Initiating Calibration Verification tests from the Quality Tests menu allows results to be stored in separate
categories for the purpose of documentation and review. Also, measurement limits are not applied to results
in this test path. Therefore, results below and above the measurement ranges will be reported.
• DO NOT insert cartridge to start test.
• DO NOT open cartridge pouch before scanning the barcode.
• If Quality Check Code 69 or 140 appears, see Analyzer Coded Messages Technical
Bulletin for Troubleshooting Information.
1. Press
to turn on handheld.
2. Press
3
MENU
3. Follow handheld prompts.
4. Scan the lot number on the cartridge pouch.
• Position barcode 3 - 9 inches from scanner window on the handheld.
• Press and hold
SCAN
• Align the red laser light so it covers the entire barcode.
• The handheld will beep when it reads the barcode successfully.
5. Continue normal procedures for preparing the sample, filling and
sealing the cartridge.
6. Push the sealed cartridge into the handheld port until it clicks into
place. Wait for the test to complete.
Note: For ACT, PT/INR, Hct, and immunoassay testing, the handheld
must remain on a level surface with the display facing up dur-
ing testing. A level surface includes running the handheld in the
downloader/recharger.
7. Review results.

Troubleshooting Cartridge Tests

See Troubleshooting Out-of-Range Results for Cartridges paragraph in the Performing Control Tests on
Cartridges section of this manual.
15-10
3
for Cal Ver Samples.
to activate the scanner.
Art: 714377-01O
Rev. Date: 27-Jan-17