2. Save the settings by selecting 9 (Save and Exit) at the Your choice ?
3. Remove power and connect the Downloader in its intended location.
Troubleshooting
If a wrong number is entered, which cannot be corrected, press the Enter key
until the session is completed and start from the beginning again.
Configure for
Channel 1 provides network access for the i-STAT 1 Analyzer data transmissions
i-STAT 1 Serial
to a data manager. This section describes how to set up parameters for Channel
Downloader
1.
Connection
1. Determine the following information:
2. At the Your choice ? prompt, Select 1 (Channel 1 Configuration).
3. At each of the prompts enter the following bold-faced value:
Art: 714383-01F
prompt.
•
The IP Address of the data manager. (Example: 10.10.12.184)
•
The service port number set to receive transmissions from the i-
STAT 1 Analyzer (default 6004).
Note: If the information to be entered is the same as the default
value, press the Enter key.
•
Baudrate (38400) ? 38400 (must be set to 38400)
•
I/F Mode (4C) ? (press <Enter> key)
•
Flow (00) ? (press <Enter> key)
•
Port No (10001) ? (press <Enter> key)
•
ConnectMode (C1) ? C1 (must be set to C1)
•
Auto increment source port (N) ? N (press <Enter> key)
•
Remote IP Address : (000)10.(000)10.(000)12.(000) 184
Note: Set Remote IP Address to the IP Address of the
computer where the data manager resides.
•
Remote Port (06004) ? 6004
Note: Remote Port refers to i-STAT 1 service port defined in
the data manager.
•
DisConnMode (00) ? (press <Enter> key)
•
FlushMode (44) ? 44 (must be set to "44")
•
DisConnTime (00:30) ? 00:30 (Disconnect time must be 30 sec-
onds)
•
SendChar 1 (00) ? (press <Enter> key)
•
SendChar 2 (00) ? (press <Enter> key)
Rev. Date: 02-Sep-08
21-5