10. Select the method in which control lot number information will be entered into the handheld.
Scan or Enter: allows the user the option of manually entering the liquid QC lot
information into the handheld, or scanning it from the barcode on the quality control vial
being tested.
Scan only: the fluid lot information must be entered by scanning the barcode on the
control vial being tested.
Note 10.1: The APOC Fluid Lot Entry Method is also customizable through the i-STAT 1
handheld keypad.
1. Power on the i-STAT 1 handheld and press MENU to get to the Administration
2. Press 4 – Customization
3. Press 2 – Change
4. Type in your password and press ENT (If no password is set, just press ENT)
5. Press 4 – QC Tests
6. Press 2 – Cartridge QC
7. Press 5 to set APOC Fluid Lot Method
11. To set up a Liquid QC Schedule, click on the Liquid QC Schedule 1 tab at the bottom of the
screen.
12. Select the frequency at which you want liquid QC to be run under this schedule.
Off: Disables the selected QC Schedule
Daily
Weekly: A particular day of the week (e.g. every Monday)
Monthly: A particular day of the month (e.g. the second Tuesday of the month).
13. Enter the QC Time. The QC Time sets the time when QC Cycles (a test run in the Control
pathway consisting of a QC cartridge and a corresponding QC fluid) will begin to count toward
satisfying the QC test profiles, i.e. when QC will become "due to start".
Note 13.1: Use the 24 hour clock designation to indicate the time when QC is due. For
example, if QC will be due at 2 pm, enter 14:00 for the time.
Rev. Date: 21-Mar-14
Menu
Note: Abbott Point of Care Inc. recommends changing the default password.
Art: 730077-01C
7