APPLY QC SCHEDULE
The months of the year to which this schedule will apply.
TO
Options are:
All months
Selected Months: Check the box next to the months to which this schedule
will apply.
CARTRIDGE QC PROFILE The System Administrator defines a QC cartridge set consisting of:
A QC cartridge type (i.e. the cartridge type to be test with specified liquid QC
fluids during the QC procedure), as well as
Any number of dependent cartridge types (i.e. associated cartridge types that
will be enabled by the handheld if the QC requirements for a given cartridge
set are met on that handheld).
The System Administrator then associates the defined cartridge set with up
to six (6) specific QC fluids.
PREFERENCE WINDOW:
for
Option
REFERENCE RANGES
Reference ranges can be defined for each test. The ranges will be
depicted as tic marks on the bar graphs on the result pages. There
are no bar graphs for blood gas, coagulation, and immunoassay
tests.
ACTION RANGES
High and low action ranges can be defined for each test.
CUSTOM REPORTABLE
High and low custom Reportable Ranges can be defined for each
RANGES
analyte (except ACT).
PRINT REFERENCE
Reference Ranges can be printed with results. Ranges will print
RANGES
only if the record to be printed is stored with the active Preference
set in the handheld.
r
r
o
esults
eportIng
ptIons
Description
All months
Disabled
A Cartridge QC Profile can only be edited or created if QC
Frequency has been activated.
See the Technical Bulletin "Liquid Quality Control Schedule
and Lockout Customization on the i-STAT 1 Handheld" for
full details on defining a Cartridge QC Profile.
Default
Ranges are listed in
Ranges will be displayed on the Customization screen of the
handheld under the Administration Menu.
the Cartridge and Test
Information sheets.
Only one range is allowed for each test in a particular handheld.
However, different customization profiles can be set up in specific
handhelds used for specific patient populations.
Care should be taken to enter the same units as selected in the Unit
Set Window.
Care should be taken to enter Action Ranges within the
Disabled
reportable ranges of the tests.
(-99999.9 to 99999.9)
Care should be taken to enter the same units as selected in
the Unit Set Window.
Disabled
See the "Reportable Range Customization on the i-STAT 1 Handheld"
Technical Bulletin for full details.
(-99999.9 to 99999.9)
Disabled
The active Preference set in the handheld is listed as "Custom" on
the Analyzer Status page and the Preference set stored with the
record is displayed on the Chart Page when the record is recalled and
is printed with the results.