Abbott i-STAT 1 System Manual page 250

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

1. 1 5
Follow the onscreen instructions.
1) If an analyzer is already in the Downloader
remove it.
2) Ensure the analyzer to be updated is off.
3) Place the analyzer in the Downloader.
*
If using the Serial Downloader/Recharger, a
If using the Serial Downloader, a
1. 1 6
When the update is in progress, the following
screen will appear:
The application update is in progress.
Please do not remove the analyzer from
the Downloader.
Programming application block 108/2055
The Receiving Handheld will have 1's and 0's streaming
across the screen signifying that it is receiving the software.
Do not move the Handheld until the success screen is displayed.
The application update was successful.
The CLEW update was successful.
Art: 731335-01B Rev. Date 13-AUG-13
Cancel
*
blue light
will illuminate when the Handheld is placed correctly within it.
red light
will illuminate when the Handheld is placed correctly within it.
Cancel
Close
If you do not see the screen
shown on the left—
—acknowledge the error
message(s) and click OK.
Return to Step
1. 1 4