PERFORMING PATIENT ANALYSIS
• DO NOT insert cartridge to start test.
• DO NOT open cartridge pouch before scanning the barcode.
• If Quality Check Code 69 or 140 appears, see Analyzer Coded Messages Technical
Bulletin for Troubleshooting Information.
1. Press
to turn on the handheld.
2. Press
for i-STAT cartridge.
2
3. Follow the handheld prompts.
4. Scan the lot number on the cartridge pouch.
• Position barcode 3-9 inches from scanner window on the
handheld.
• Press and hold
• Align the red laser light so it covers the entire barcode.
• The handheld will beep when it reads the barcode
successfully.
5. Continue normal procedures for preparing the sample and
filling and sealing the cartridge.
6. Push the sealed cartridge into the handheld port until it clicks
into place. Wait for the test to complete.
Note: For ACT, PT/INR, Hct, and immunoassay testing, the
handheld must remain on a level surface with the display
facing up during testing. A level surface includes running
the handheld in the downloader/recharger.
7. Review results.
12-2
SCAN
to activate the scanner.
Art: 714374-01M
Rev. Date: 23-Feb-16