Note 7.2: The Control Test Settings feature is also customizable through the i-STAT 1 handheld
keypad.
1. Power on the i-STAT 1 handheld and press MENU to get to the Administration
2. Press 4 – Customization
3. Press 2 – Change
4. Type in your password and press ENT (If no password is set, just press ENT)
5. Press 4 – QC Tests
6. Press 2 – Cartridge QC
7. Press 2 to set Comment Codes for in range results or press 3 to set Comment
8. Select the way in which you would like control results to be displayed.
Numeric: liquid QC results are displayed in numeric format
Suppressed: the following symbol "< >" is displayed next to each liquid QC test name
in place of the quantitative (numeric) results.
Note 8.1: The "Suppressed" option should only be chosen if "Auto via eVAS" is chosen
for the liquid QC Pass/Fail Determination.
Note 8.2: The Control Results Display Format feature is also customizable through the
i-STAT 1 handheld keypad.
9. Select the method in which control lot number information will be entered into the handheld.
Scan or Enter: allows the user the option of manually entering the liquid QC lot
information into the handheld, or scanning it from the barcode on the quality control vial
being tested.
Scan only: the fluid lot information must be entered by scanning the barcode on the
control vial being tested.
Rev. Date: 21-Mar-14
Menu
Note: Abbott Point of Care Inc. recommends changing the default password.
Codes for out of range results
1. Power on the i-STAT 1 handheld and press MENU to get to the
Administration Menu
2. Press 4 – Customization
3. Press 2 – Change
4. Type in your password and press ENT (If no password is set, just press ENT)
Note: Abbott Point of Care Inc. recommends changing the default
password.
5. Press 4 – QC Tests
6. Press 2 – Cartridge QC
7. Press 4 to set Result Format
Art: 730077-01C
13