Suppressed Control Result Display with a Passed QC run: All analyte results will display a <>
symbol and the overall Pass assessment appears at the top of the display screen above the
results.
Suppressed Control Result Display with a Failed QC run: All analyte results will display a <>
symbol and the overall Fail assessment appears at the top of the display screen above the
results. The status of individual failed results is displayed on the handheld as either an up-arrow
(
) for out-of-range high results, or a down-arrow (
h
HANDHELD DISPLAY MESSAGES
The following messages may appear on the i-STAT 1 handheld when it is customized for Auto QC P/F:
Message
Invalid eVAS
Lot Not in eVAS
Does Not Match Selected Level
Invalid Length
Lot Expired
Invalid Number
Rev. Date: 29-Jul-14
) for out-of-range low results.
i
The eVAS file does not exist or is invalid.
Download current eVAS file from the APOC
website Value Assignment Sheet page.
The entered cartridge or control lot number
could not be found in eVAS file. Download
current eVas file from the APOC website Value
Assignment Sheet page.
The entered control solution lot number does
not match the user selected control level.
A blank control solution lot number was
The entered control solution lot number or
Cartridge lot number is expired.
The scanned APOC lot number has an invalid
Art: 730078-01C
Cause
entered.
format.