Clicking on an individual alarm category opens a window displaying specific details of the alarms detected
in that category, the date/time they occurred, and in what location.
Selected Alarm details or all Alarm details may be deleted by clicking the Acknowledge button at the bottom
of the window.
A report of selected Alarm details or all Alarm details may be printed by clicking the Print button at the
bottom of the window.
The default setting for the Alarm Monitor is to have all Alarm Categories and the individual detailed alarms
contained in those categories enabled. To disable alarm reporting, alarm categories, or specific alarms
contained in those categories:
1. Close the CDS application, if open.
2. Click start → run...
3. Type wcds32 config at the open: prompt and click oK.
4. When the i-STAT Central Data Station Customization window appears, click on the Alarms tab.
5. To disable alarm reporting, uncheck the "Enable alarm reporting" box and click Accept.