Abbott i-STAT 1 System Manual page 130

Hide thumbs Also See for i-STAT 1:
Table of Contents

Advertisement

Note 9.1: The APOC Fluid Lot Entry Method is also customizable through the i-STAT 1 handheld
keypad.
1. Power on the i-STAT 1 handheld and press MENU to get to the Administration
2. Press 4 – Customization
3. Press 2 – Change
4. Type in your password and press ENT (If no password is set, just press ENT)
5. Press 4 – QC Tests
6. Press 2 – Cartridge QC
7. Press 5 to set APOC Fluid Lot Method
10. To set up a Liquid QC Schedule, click on the Liquid QC Schedule 1 tab at the top of the screen.
11. Select the frequency at which you want liquid QC to be run under this schedule.
Off: Disables the selected QC Schedule
Daily
A particular day of the week (e.g. every Monday)
A particular day of the month (e.g. the second Tuesday of the month).
12. Select the months of the year in which you want this schedule to apply.
All months
Selected months. Check the box next to the months to which you want this schedule to
apply.
13. Enter the QC Time. The QC Time sets the time when QC Cycles (a test run in the Control
pathway consisting of a QC cartridge and a corresponding QC fluid) will begin to count toward
satisfying the QC test profiles, i.e. when QC will become "due to start".
Note 13.1: Use the 24 hour clock designation to indicate the time when QC is due. For
example, if QC will be due at 2 pm, enter 14:00 for the time.
14
Menu
Note: Abbott Point of Care Inc. recommends changing the default password.
Art: 730077-01C
Rev. Date: 21-Mar-14