Downloader/
Recharger Indicator
LEDs

Power Requirements

The Downloaders require one power outlet. The Downloader and Downloader/
Recharger must be used with the AC power supply adapter supplied with them.
The Downloaders are capable of supplying power to the portable printer which
reduces the number of power outlets required in the downloading and printing
area.
DR Affect on
The operating temperature for an i-STAT 1 Analyzer is 16°C to 30°C. The DR
Ambient Operating
and Rechargeable Battery may raise the temperature of the i-STAT 1 Analyzer
Temperature Range
2°C-3°C relative to the ambient temperature if:
Programming and
Details for programming the Network Downloaders can be found in the
Connections
Downloader Programming and Wiring section of this manual. Diagrams and
instructions for connecting peripheral components to the Downloader can also
be found in the Downloader Programming and Wiring section.

Cautions

The Downloader and Downloader/Recharger are not intended for use in the
patient environment (within 1.5 meters of the physical
location of the patient).
Users should not connect the Downloader or the
Downloader/Recharger to a medical electrical system.
Do not place metal objects on or near the exposed gold
charging contacts.
Be sure to install all cables and power supplies so they
do not pose a trip hazard. Mount equipment so cables
and accessories stay clear of walkways. The AC power
supply adapter plug acts as the disconnect device for the
Downloader and Downloader/Recharger and, therefore,
the socket outlet must be installed (or located) near
the Downloader or Downloader/Recharger and must
be easily accessible.
Only i-STAT provided printers may be connected to the
Downloader printer port.
An ethernet cable and serial (DB9) cable may NOT be connected to the
Downloader at the same time.
Art: 714368-01J
Analyzer Battery LED (near top of Downloader/Recharger)
Off
Blinking Red
Solid Red
Solid Green
S
B
(near middle of Downloader/Recharger)
pare
attery
Off
Green
The Analyzer is frequently lifted and replaced into the DR
Multiple thermally controlled cartridges are run in the Analyzer
while it is in the DR.
Rev. Date: 02-Aug-12
No Rechargeable Battery
Fast Charge Pending
Fast Charging
Trickle Charging
No Rechargeable Battery
Trickle Charging